Application fields
Hidex AMG GammaDx
Hidex AMG GammaDx is an automated in vitro diagnostic gamma counter designed for the quantitative measurement of gamma radiation emitted from human specimens. GammaDx is intended to help nuclear medicine and clinical laboratory professionals achieve reliable measurement performance and efficient routine operation.
Reliable gamma counting for IVD applications
- Automated gamma counting for professional IVD laboratory use
- Designed for nuclear medicine and clinical laboratory workflows
- Automated handling of multiple sample racks
- Integrated quality control, background monitoring and Cs-137 calibration
- Automatic decay and background correction
GammaDx for nuclear medicine and clinical laboratories
GammaDx supports professional nuclear medicine and clinical laboratory workflows in which serum, plasma, tissue extracts or other human specimens are prepared using commercially available IVD kits containing gamma-emitting tracers where its use is consistent with the kit instructions containing gamma-emitting tracers. The instrument provides quantitative radiation measurement data to support diagnostic or therapeutic decision-making.
With automated sample handling, flexible measurement templates and integrated quality-control functions, GammaDx helps laboratories achieve an efficient, traceable and controlled gamma-counting workflow.
Controlled workflow for patient-associated specimens
For clinical use, assay settings and measurement templates are selected and controlled according to the applicable IVD kit instructions, the validated examination procedure and the laboratory quality-management system.
Optional features:
Sample Balance
2D vial ID reader providing additional sample identification and traceability
21 CFR part 11 security support software
Extended MCA for high energy isotopes
Technical data
|
Operational sample capacity |
Ø13mm max 230 pcs. and Ø 28 mm max 72 pcs |
|
Detector type |
3 inch well type NaI detector |
|
Counting efficiency (isotope peak window) |
¹³⁷Cs 18%, ¹²⁹I 58% |
|
Energy range |
15-2047 keV |
|
Lead shield thickness |
55 mm with up to 80 mm on conveyor side |
|
Dimensions, W/D/H (cm) |
62/70/60 with balance 62/83/60 |
|
Weight |
198 kg (w/o balance) |
|
Power requirement |
100-240 V / 50-60 Hz |
|
Operating system for user interface: |
Microsoft Windows 11 |
|
Instrument connection: |
USB communication |
Product availability may vary by country. Please contact Hidex or your local distributor for information about availability in your market.